Calibrated channel number. This parameter is ignored; the verification is either
enabled or disabled for all channels.
Calibrated port number. This parameter is ignored; the verification is either
enabled or disabled for all sources.
Enables (ON) or disables (OFF) the verification sweep.
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire],
SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]:VERification:RESult?
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage[:COUNt] <readings>
Sets a limit for the number of calibration sweeps in the source power calibration. The command is valid for
all channels, ports and external generators. Equivalent command:
SOURce<Ch>:POWer<Pt>:CORRection:NREadings.
Calibrated channel number. This parameter is ignored; the limit applies to all
channels.
Calibrated port number. This parameter is ignored; the limit applies to all
sources.
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire] (for equivalent command
SOURce<Ch>:POWer<Pt>:CORRection:NREadings).
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage:NTOLerance
<tolerance>
Specifies the maximum deviation of the measured power from the target power of the calibration. The
command is valid for all channels and calibrated ports.
Calibrated channel number. This parameter is ignored; the tolerance value is
valid for all sources.
Calibrated port number. This parameter is ignored; the tolerance value is valid
for all sources.
Device-specific, command or query