See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:CFACtor <convergence>
Specifies the convergence factor for a source power calibration.
Calibrated channel number. This parameter is ignored; the convergence
factor is valid for all sources.
Calibrated port number. This parameter is ignored; the convergence factor is
valid for all sources.
0 to 2 [–]. UP and DOWN increment/decrement the convergence factor in 1-
dB steps.
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness <Boolean>
Enables or disables the source power calibration. The command is valid for all channels, ports and
external generators.
It is not possible to disable flatness calibration and the reference receiver calibration simultaneously
(SOURce<Ch>:POWer<Pt>:CORRection:COLLect:RRECeiver).
Calibrated channel number. This parameter is ignored; the verification is either
enabled or disabled for all channels.
Calibrated port number. This parameter is ignored; the verification is either enabled
or disabled for all sources.
Enables (ON) or disables (OFF) the flatness calibration. With disabled flatness
calibration, only one calibration sweep is performed in order to calibrate the
reference receiver; the previous source calibration data is not overwritten.
ON
Flatness calibration is automatically disabled when a frequency converter setup
without electronic attenuator is selected in the System Configuration dialog.
Device-specific, command or query
*RST; SOUR:POW:CORR:COLL:VER 0
Disable the verification sweep to speed up the source power calibration procedure.
SOUR:POW:CORR:COLL:FLAT OFF; RREC ON