Operating Manual 1145.1084.12 – 30 1018
Disable flatness cal and enable reference receiver calibration.
SOUR:POW:CORR:ACQ PORT,1
Perform a single source power calibration sweep using port 1 as a source port.
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:METHod PMONly | RRAFter |
RRONly
Selects the source power calibration method.
Calibrated channel number. This parameter is ignored; the method is valid for all
channels.
Calibrated port number. This parameter is ignored; the method is valid for all sources.
PMONly – Power Meter only (requires
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness ON and
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:RRECeiver OFF)
RRAFter – Reference Receiver after the number of readings specified via
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:PMReadings (requires
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness ON)
RRONly – Reference Receiver only (requires
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:RRECeiver OFF)
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:PMReadings <no_readings>
Selects the number of power meter readings for source power calibration method
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:METHod RRAFter.
Calibrated channel number. This parameter is ignored; the parameter is valid for
all channels.
Calibrated port number. This parameter is ignored; the parameter is valid for all
sources.
Number of power meter readings. If the value exceeds the maximum number of
calibration sweeps
(SOURce<Ch>:POWer<Pt>:CORRection:COLLect:AVERage[:COUNt]) the latter
is adjusted.
Device-specific, command or query