R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1019
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:COLLect:RRECeiver <Boolean>
Enables or disables calibration of the reference receiver together with the source power calibration.
It is not possible to disable flatness calibration
(SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness) and the reference receiver calibration
simultaneously.
Calibrated channel number. This parameter is ignored; the verification is either
enabled or disabled for all channels.
Calibrated port number. This parameter is ignored; the verification is either
enabled or disabled for all sources.
Enables (ON) or disables (OFF) reference receiver calibration.
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection:COLLect:FLATness
SOURce<Ch>:POWer<Pt>:CORRection:CONVerter<Con>:LEVel:OFFSet <offset>
Specifies a gain (positive values) or an attenuation (negative values) in the signal path between the
converter source port <Con> and the calibrated reference plane (Cal Power Offset). The value has no
impact on the source power.
Calibrated channel number
Calibrated port number. This suffix is ignored; the converter is selected via
<Con>.
Gain or attenuation value
Device-specific, command or query
See SOURce<Ch>:FREQuency<Pt>:CONVersion:ARBitrary:CFRequency
SOURce<Ch>:POWer<Pt>:CORRection:CONVerter<Con>[:STATe] <Boolean>
Enables or disables the source power calibration for channel <Ch> and for an external frequency
converter number <Con>. The command is disabled unless a source power calibration for the external
converter has been performed (SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire] CONVerter,
<source_pt>). To enable or disable a source power calibration for an analyzer port use