R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1020
SOURce<Ch>:POWer<Pt>:CORRection[:STATe].
Calibrated channel number
Calibrated port number. This suffix is ignored; the converter is selected via
<Con>.
Enables (ON) or disables (OFF) the source power calibration for the
converter <Con>.
Device-specific, command or query
See SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire].
SOURce<Ch>:POWer<Pt>:CORRection:DATA '<src_string>', <block> |
<numeric_value>...
Reads or writes source power correction data sets. A power correction data set contains n real values
where:
Each value corresponds to the ratio of the actual power at the reference plane (value provided by
the used source) to the uncalibrated power in dB.
The number n is equal to the number of sweep points.
Increasing (decreasing) the values in the correction data sets increases (decreases) the power at the
reference plane. Writing correction data (the setting command) fails if the number of transferred values is
not equal to the number of sweep points.
Calibrated channel number
Calibrated port number. This parameter is ignored; the port is identified
by the '<src_string>'.
Identifier for the source of the calibrated wave:
'A<n>' or 'PORT<n>' denote correction data for the analyzer port
no. <n>.
'ESRC<n>' or 'GEN<n>' denote correction data for the external
generator no. <n>.
'CONV<n>' denotes correction data for a frequency converter
(e.g. R&S ZVA-Z110E) at port no. <n>
<block> |
<numeric_value>...
Power correction values in block data format or as a comma-separated
list of real numbers. The query returns a list of numbers.
All numbers are interpreted as dB-values.
Device-specific, command or query
*RST; :SWE:POIN 10
Reset the instrument and reduce the number of sweep points to 10.
SOUR:POW:CORR:ACQ PORT,1