R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1067
– (neither the password nor the protection of service functions is affected by
*RST)
SYST:PASS "XXXX"
Enter password.
SYSTem:PRESet
Presets the active setup. Restores factory-defined or user-defined default settings (see
SYSTem:PRESet:REMote[:STATe]).
Other open setups remain unchanged.
SYST:PRES
Reset the parameters of the current setup.
SYSTem:PRESet:REMote[:STATe] <Boolean>
Defines whether the remote commands *RST and SYSTem:PRESet shall use factory-defined or user-
defined default settings (see PRESet:USER[:STATe]).
OFF - always restore factory-defined settings (Factory Preset)
ON - restore user-defined settings
– (*RST does not affect the preset state setting)
Device-specific, command or query
See SYSTem:PRESet:USER[:STATe]
SYSTem:PRESet:SCOPe ALL | SINGle
Defines whether a manual Preset (via GUI or front panel key) shall affect all open setups or the active
setup only.
On the R&S ZVA, a SYSTem:PRESet:SCOPe ALL automatically executes a SYSTem:PRESet.
All open setups are closed and a single setup "Set1" is created with default
settings
The settings of the active setup are reset; the name of the active setup and the
parameters of all other setups remain unchanged
– (*RST does not affect the preset scope setting)