ZVT GUI Reference
Channel Menu
Operating Manual 1145.1084.12 – 30 344
shows the wave quantity b
2
. It is possible, however, to display other quantities and change the trace
settings. A label H2 in the channel list indicates that the second harmonic measurement is active.
In a frequency sweep the 2
nd
harmonic is displayed as a function of the stimulus frequency.
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:ORDer 2
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt 2
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:SPORt 1
Alternative: [SENSe<Ch>:]FREQuency:CONVersion SHARmonic
3rd Harmonic Port1 to Port2
Activates the measurement of the third harmonic at port 2 with port 1 as source port. The active trace
shows the wave quantity b
2
. It is possible, however, to display other quantities and change the trace
settings. A label H3 in the channel list indicates that the third harmonic measurement is active.
In a frequency sweep the 3rd harmonic is displayed as a function of the stimulus frequency.
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:ORDer 3
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt 2
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:SPORt 1
[SENSe<Ch>:]FREQuency:CONVersion THARmonic
More Harmonics...
Opens a dialog for higher harmonics and arbitrary source and receive ports.
The Source port provides the stimulus signal at the fundamental frequency.
The receive port (Harmonics Measured at:) measures the signal at n times the fundamental
frequency where n is the Harmonic Order.
The Harmonic Order can be set in the range between 2 (2nd harmonic) and 100.
A label H<n> in the channel list indicates that the nth harmonic measurement is active.
In a frequency sweep all harmonics are displayed as a function of the stimulus frequency.