R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 633
CALCulate<Chn>:LIMit:RDOMain:COMPlex S | SINV | Y | Z | YREL | ZREL
[Deprecated command, not needed on R&S ZVA]
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via CALCulate<Chn>:LIMit:CONTrol:DOMain.
This command is complemented by CALCulate<Chn>:LIMit:RDOMain:FORMat and
CALCulate<Chn>:LIMit:RDOMain:SPACing.
Channel number used to identify the active trace.
Keyword for the physical unit of the response values.
The parameters form four groups:
S and SINV select relative units (dB) for the limit line.
Y selects admittance units (S/Siemens).
Z selects impedance units (Ω).
YREL and ZREL select dimensionless numbers (U).
[–]
Device-specific, no query.
CALCulate<Chn>:LIMit:RDOMain:FORMat COMPlex | MAGNitude | PHASe | REAL |
IMAGinary | SWR | GDELay | L | C
[Deprecated command, not needed on R&S ZVA]
Deletes the existing limit line and (re-)defines the physical units of the response values of the limit line.
The units of the stimulus values are defined via CALCulate<Chn>:LIMit:CONTrol:DOMain.
This command is complemented by CALCulate<Chn>:LIMit:RDOMain:COMPlex and
CALCulate<Chn>:LIMit:RDOMain:SPACing.
Channel number used to identify the active trace.
Keyword for the physical unit of the response values.
The parameters form six groups:
COMPlex, REAL, IMAGinary, and SWR select dimensionless
numbers (U) for the limit line.
MAGNitude selects relative units (dB).
PHASe selects phase units (deg).
GDELay selects time units (s).
L selects inductance units (H/Henry).