Operating Manual 1145.1084.12 – 30 641
CALCulate<Chn>:LIMit:UPPer:STATe <Boolean>
Switches the upper limit check on or off. Upper limit line segments are assigned odd numbers; see
CALCulate<Chn>:LIMit:UPPer[:DATA]. The command does not affect segments with even
numbers.
Use CALCulate<Chn>:LIMit:STATe to switch on or off the entire limit check, including upper and
lower limit lines.
Channel number used to identify the active trace.
ON | OFF - Limit check on or off.
Confirmed, command or query.
CALC:LIM:UPP -10, 0, 0, -10
Define the following upper and (default) lower limit line segments:
CALC:LIM:UPP:STAT ON; :CALC:LIM:FAIL?
Switch the limit check on and query the result.
CALCulate<Chn>:MARKer<Mk>...
This subsystem controls the marker functions. The commands are device-specific and beyond what is
specified in the SCPI subsystem SOURce:MARKer.
CALCulate<Chn>:MARKer<Mk>:AOFF
Removes all markers from all traces of the active setup. The removed markers remember their properties
(stimulus value, format, delta mode, number) when they are restored (CALC<Chn>:MARK<Mk> ON). The
marker properties are definitely lost if the associated trace is deleted.
Channel number used to identify the active trace. If unspecified the numeric
suffix is set to 1.
Marker number in the range 1 to 10. This numeric suffix is ignored and may
be set to any value.
Device-specific, no query.
Suppose that the active setup contains an active trace no. 1.
CALC:MARK1 ON; MARK2 ON