Channel number used to identify the active trace.
ON | OFF - Fail beep on or off.
Confirmed, command or query.
CALC:LIM:STAT ON; SOUN ON
Switch the limit check on and activate the fail beep.
CALCulate<Chn>:LIMit:STATe
Switches the limit check (including upper and lower limits) on or off.
Use CALCulate<Chn>:LIMit:UPPer:STATe or CALCulate<Chn>:LIMit:LOWer:STATe to
switch on or off the individual limit checks for upper or lower limit lines.
Channel number used to identify the active trace.
ON | OFF - Limit check on or off.
Confirmed, command or query.
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between 1 GHz and 2
GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALCulate<Chn>:LIMit:TTLout<Output_no>[:STATe] <Boolean>
Switches the TTL pass/fail signals on or off. The signals are applied to the USER CONTROL connector as
long as the active trace <Chn> is within limits, including the ripple limits.
Channel number used to identify the active trace.
1 – TTL out pass 1 (pin 13 of USER CONTROL connector)
2 – TTL out pass 2 (pin 14 of USER CONTROL connector)
ON | OFF - TTL output signal on or off.
Device-specific, command or query.
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range between 1 GHz and 2
GHz, using default response values.