Query whether all data traces are displayed. The response 0 means that at
least one trace is hidden.
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:X:OFFSet <numeric_value>
Shifts the trace <WndTr> in horizontal direction, leaving the positions of all markers unchanged.
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON).
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED.
–1000 GHz to +1000 GHz [Hz, for frequency sweeps]. The range and unit
depends on the sweep type ([SENSe<Ch>:]SWEep:TYPE).
Device-specific, with query.
*RST; :DISP:WIND:TRAC:X:OFFS 1MHZ; :DISP:WIND:TRAC:Y:OFFS 10
Create the default trace and shift it horizontally by 1 MHz, vertically by 10 dB.
DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:Y:OFFSet <Magnitude>[,<Phase>,
<Real>, <Imaginary>]
Modifies all points of the trace <WndTr> by means of an added and/or a multiplied complex constant. The
response values M of the trace are transformed according to:
Number of an existing diagram area (defined by means of
DISPlay:WINDow<Wnd>:STATe ON).
Existing trace number, assigned by means of
DISPlay:WINDow<Wnd>:TRACe<WndTr>:FEED.
Multiplied magnitude factor
Multiplied phase factor, optional for setting command but returned by query
–3.4*10
38
deg to +3.4*10
38
deg [deg]
Real and imaginary part of added complex constant, optional for setting
command but returned by query
–3.4*10
38
to +3.4*10
38
[–]