Operating Manual 1145.1084.12 – 30 815
Since FW version 3.40 this is a global setting.
Channel number. This suffix is ignored, the command affects all channels.
See Port<i> Power Control
NONE - None or Mechanical Attenuator
ELECtronic - Electronic Attenuator only
LAPProx - RF IN power (and Electronic Attenuator) with Linear
Approximation
DSET - RF IN power (and Electronic Attenuator) with Data Set
n/a (a *RST does not affect the general frequency converter settings)
Device-specific, command or query
See [SENSe<Ch>:]CONVerter:AMODel
[SENSe<Ch>:]CONVerter:PATH '<directory>'
[Deprecated command, superseded by SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:PATH]
Selects the data directory for the RF In Power (and Electronic Attenuator) with Data Set power transfer
model for all converter ports.
Since FW version 3.40 this is a global setting.
Channel number. This suffix is ignored, the command affects all channels.
String parameter to specify an existing directory path.
n/a (a *RST does not affect the general frequency converter settings)
Device-specific, command or query
CONV:AMOD ON; DESC DSET
Enable the advanced power transfer model based on a data set for all converter
ports.
CONV:PATH 'C:\Program Files\Rohde&Schwarz\Network Analyzer\FrequencyConverter'
Define a directory for calibration files. 'C:\Program
Files\Rohde&Schwarz\Network Analyzer\FrequencyConverter' is the factory
default setting.
[SENSe<Ch>:]CORRection...
This subsystem controls system error correction and recording of correction data.
[SENSe<Ch>:]CORRection:CBFReq[:STATe] <Boolean>
Forces system error corrections to be performed at the channel base frequency f
b
, irrespective of the port