Device-specific, command or query
See One and Two-Port Calibration
[SENSe<Ch>:]CORRection:FACTory[:STATe] <Boolean>
Enables or disables the factory calibration for all channels in the active setup.
Channel number. This parameter is ignored; the command affects all
channels.
Enable or disable factory calibration
Device-specific, command or query
*RST; :CORR:FACT?
Reset the instrument and query whether the factory calibration for channel 1 is
enabled. The response is 1.
[SENSe<Ch>:]CORRection:LOSS<port_no> <ref_loss>
Defines the offset loss at the reference frequency
([SENSe<Ch>:]CORRection:LOSS<port_no>:FREQuency).
Channel number of the offset-corrected channel
Port number of the analyzer
Frequency-dependent part of the offset loss
-200 dB to +200 dB [dB]. The increment (UP/DOWN) is 0.001 dB.
Device-specific, command or query
See [SENSe<Ch>:]CORRection:EDELay<port_no>:ELENgth
[SENSe<Ch>:]CORRection:LOSS<port_no>:AUTO ONCE
Defines the offset parameters for the active test port such that the residual delay of the active trace
(defined as the negative derivative of the phase response) is minimized and the measured loss is
reproduced as far as possible across the entire sweep range.
Channel number of the offset-corrected channel.
Port number of the analyzer. This numeric suffix is ignored; the active port is
determined by the active trace.
Applies the Auto Length and Loss function.