[SENSe<Ch>:]CORRection:STIMulus?
Queries the stimulus values of the active calibration. A calibration must be selected before the command
is executed; see example.
Channel number of the calibrated channel.
Device-specific, query only
*RST; :CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as calibration
type.
CORR:STIM?
Query the stimulus frequencies. The response contains 201 frequency values.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the measurement
results of this standard.
CORR:COLL:SAVE:SEL
Calculate the system error correction data and apply it to the active channel.
[SENSe<Ch>:]COUPle...
This subsystem selects the sweep type and the measurement parameter.
[SENSe<Ch>:]COUPle ALL | NONE
Activates normal or alternating sweep mode.
Note that some measurement modes are not compatible with alternate sweep mode (see GUI descritpion
for details). While one of these modes is active, the analyzer firmware automatically switches to normal
("chopped") mode.
Channel number. If unspecified, this parameter is set to 1.
ALL – normal sweep mode (alternate off), complete all partial measurements
before proceeding to the next sweep point
NONE – alternating sweep mode on, reverse the order of partial measurements
and sweeps
Device-specific, command or query