R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 928
Number of the pulse generator signal. This suffix must be set to 1, since the
command affects the pulse generator signal only.
ON | OFF - Segment active or inactive
ON (no pulse train segment defined after *RST)
Device-specific, command or query.
See [SENSe<Ch>:]PULSe:GENerator<gen_no>:TRAin:DATA
[SENSe<Ch>:]PULSe:GENerator<gen_no>:TRAin:SEGMent<Seg>:STOP <Stop
High>
Changes the stop stimulus value (i.e. the largest stimulus value) of a pulse train segment. A range must
be created first to enable this command (see
[SENSe<Ch>:]PULSe:GENerator<gen_no>:TRAin:DATA).
Number of the pulse generator signal. This suffix must be set to 1, since the
command affects the pulse generator signal only.
12.5 ns to 54975.5813632 s [s]. The minimum width of a pulse is 12.5 ns, its
maximum width is given by the pulse train period
([SENSe<Ch>:]PULSe:GENerator<gen_no>:TRAin:PERiod).
12.5 ns (no pulse train segment defined after *RST)
Device-specific, with query.
See [SENSe<Ch>:]PULSe:GENerator<gen_no>:TRAin:DATA
[SENSe<Ch>:]PULSe:GENerator<gen_no>:TYPE SINGle | TRAin | CHIGh | CLOW
Selects the pulse type of the pulse generator signal <gen_no>.
Number of the pulse generator signal: 1 for pulse generator, 2 for sync
generator signal.
SINGle – Single pulse
CHIGh – Constant high
CLOW – Constant low
TRAin – Pulse train (available for pulse generator signal only, <gen_no> = 1)