R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 944
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol <Boolean>
Selects common or independent Selectivity settings for the sweep segments.
Sweep segment number. This suffix is ignored; the setting controls the whole
segmented sweep.
ON – Use independent selectivity settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect.
OFF – Reset the selectivity in all sweep segments to the selectivity for unsegmented
sweeps, defined via SENSe<Ch>:BWIDth[:RESolution]:SELect. ON will not
restore the previous values.
OFF. The parameter is automatically switched to ON when a selectivity is entered
using [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect.
Device-specific, command or query
*RST; :SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings and thus
NORMal selectivity.
SEGM:BWID:SEL HIGH
Change the selectivity to HIGH.
BWID:SEL?
Query the (default) selectivity for unsegmented sweeps. The response is NORM.
SEGM:BWID:SEL:CONT OFF
Couple the selectivities in all segments and reset the selectivity in segment no. 1 to
the unsegmented value NORMal.
[SENSe<Ch>:]SEGMent<Seg>:CLEar
[Deprecated command]
Deletes all sweep segments in the channel. The command is equivalent to
[SENSe<Ch>:]SEGMent<:Seg>:DELete:ALL.
[SENSe<Ch>:]SEGMent<Seg>:COUNt?
Returns the number of sweep segments in the channel including all segments that are switched off
([SENSe<Ch>:]SEGMent<Seg>[:STATe] OFF).
Sweep segment number. This suffix is ignored; the command counts all
segments.