:SYSTem:TIME:ZONE:CATalog?
Querys the list of available timezones.
Return values:
<Catalog>
Example: See Example"Configuring date and time" on page 519.
Usage: Query only
Manual operation: See "Timezone" on page 593
:SYSTem:TIME:PROTocol <TimeProtocol>
Sets the date and time of the operating system.
Parameters:
<TimeProtocol> OFF | NONE | 0 | NTP | ON | 1
NONE
Sets the date and time according to the selected timezone,
see :SYSTem:TIME:ZONE:CATalog? on page 540 and :
SYSTem:TIME:ZONE on page 539.
NTP
Sets the date and time derived from the network time protocol.
To select the NTP time server, use the commands :SYSTem:
NTP:HOSTname on page 539 and :SYSTem:NTP:STATe
on page 539.
*RST: n.a. (factory preset: NONE)
Example: See Example"Configuring date and time" on page 519.
Manual operation: See "Time Protocol" on page 593
:SYSTem:UPTime?
Queries the up time of the operating system.
Return values:
<UpTime> "<ddd.hh:mm:ss>"
Example:
SYSTem:UPTime?
Response: "0.08:11:00"
Usage: Query only
:SYSTem:BIOS:VERSion?
Queries the BIOS version of the instrument.
Return values:
<Version> string