Remote control commands
R&S
®
SMB100B
548User Manual 1178.3711.02 ─ 08
Table 12-1: Cross-reference between the manual and remote control
R&S proprietary value name SCPI conform value name Parameter in manual control
AUTO IMMediate "Auto" mode
SINGle BUS "Single" mode.
EXTernal EXTernal "Ext Single" and "Ext Step" mode.
Use command LFO:SWEep:MODE
to select between the two sweep
modes.
EAUTo - "Ext Start/Stop" mode.
:TRIGger<hw>:FSWeep:SOURce....................................................................................548
:TRIGger<hw>:LFFSweep:SOURce.................................................................................548
:TRIGger<hw>:PSWeep:SOURce....................................................................................548
:TRIGger<hw>[:SWEep]:SOURce....................................................................................548
:TRIGger<hw>:FPSWeep:SOURce..................................................................................549
:TRIGger<hw>:FSWeep[:IMMediate]................................................................................550
:TRIGger<hw>:LFFSweep:IMMediate.............................................................................. 550
:TRIGger<hw>:PSWeep[:IMMediate]................................................................................550
:TRIGger<hw>[:SWEep][:IMMediate]................................................................................550
:TRIGger<hw>:LFFSweep.............................................................................................. 550
:TRIGger<hw>:FSWeep:SOURce <Source>
:TRIGger<hw>:LFFSweep:SOURce <Source>
:TRIGger<hw>:PSWeep:SOURce <Source>
:TRIGger<hw>[:SWEep]:SOURce <Source>
Selects the trigger source for the corresponding sweeps:
●
FSWeep - RF frequency
●
LFFSweep - LF frequency
●
PSWeep - RF level
●
SWEep - all sweeps
The source names of the parameters correspond to the values provided in manual
control of the instrument. They differ from the SCPI-compliant names, but the instru-
ment accepts both variants.
Use the SCPI name, if compatibility is an important issue. Find the corresponding
SCPI-compliant commands in Cross-reference between the manual and remote con-
trol.
Setting parameters:
<Source> AUTO | IMMediate | SINGle | BUS | EXTernal | EAUTo
AUTO [IMMediate]
Executes a sweep automatically.
In this free-running mode, the trigger condition is met continu-
ously. I.e. when a sweep is completed, the next one starts imme-
diately.
TRIGger subsystem