EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 226

Rohde & Schwarz SMBV100A
911 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
211Operating Manual 1176.8016.02 ─ 17
Access:
Select "RF/A Mod > config... > RF Frequency > Frequency/Phase".
In the upper section of the combined "RF Frequency / Phase ..." settings dialog,
you can configure the frequency of the RF signal.
The remote commands required to define the settings are described in Chapter 7.16.7,
"SOURce:FREQuency Subsystem", on page 734.
Frequency
Sets the RF frequency of the RF output connector. The frequency entered and dis-
played here corresponds to the frequency at the RF output, that means any offset entry
is not considered.
Note: Suppressed values in the status bar
For security concerns or certain operating modes, you can hide the frequency and
level display in the status bar.
The display has been disabled for security reasons.
See:
Annotation Frequency
Annotation Amplitude
The display is disabled when list mode is running, see " State - List Mode "
on page 289.
Remote command:
[:SOURce<hw>]:FREQuency[:CW|FIXed] on page 736
Note: This command sets the frequency of the "FREQ" display, that is the frequency
containing offset.
Offset
Sets the frequency offset relative to the RF frequency. The frequency offset of a down-
stream instrument (for example a mixer) is entered.
The entry does not change the value of the RF frequency at the RF output. It only
changes the RF frequency displayed in the display header. The value of the RF fre-
quency in the header corresponds to the frequency at the output of the downstream
instrument, see also Chapter 5.5.2.1, "RF Frequency vs. RF Output Frequency",
on page 209.
Remote command:
[:SOURce<hw>]:FREQuency:OFFSet on page 739