Instrument Function
R&S
®
SMBV100A
235Operating Manual 1176.8016.02 ─ 17
"Edit"
Calls a selection of possible actions described below.
"Insert Row"
Insert a new row before the marked row.
"Insert Range"
Insert new rows before the marked row. The number of rows to be
inserted can be defined in an entry window.
"Fill...."
Opens a sub menu for defining a set of list values to be automatically
entered in the ucor list (see Chapter 5.5.9.2, "Filling the Correction
List automatically", on page 237).
"Fill With Sen-
sor"
Calls the menu to activate the filling of the user correction list with
level values acquired by the selected power sensor (see Chap-
ter 5.5.9.3, "Filling the Correction List with Power Sensor Measure-
ment Data", on page 238).
"Delete Row"
Deletes the marked row.
"Delete
Range..."
Allows to delete any number of rows starting with the marked row.
The number of rows to be deleted can be defined in an entry window.
"Save as"
Open the file menu to save the list under a new name.
Note: Save list only after filling both columns (frequency and level),
otherwise the entries are lost.
Each list is saved to the R&S SMBV hard disk as a separate file with
the file prefix *.uco. The file name and the directory to which the file
is saved are user-selectable.
"Save"
The list is saved under its current name.
Remote command:
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 729
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency on page 723
[:SOURce<hw>]:CORRection:CSET:DATA:POWer on page 724
Import/Export
User correction list can be imported from externally created files or exported into text or
CSV-files. The import/export settings are available after clicking the "Import/Export"
button.
RF Block