Remote Control Commands
R&S
®
SMBV100A
723Operating Manual 1176.8016.02 ─ 17
with correction values for external test assemblies can be automatically determined,
e.g. for compensating the frequency response of cables. The correction values can be
acquired any time irrespective of the modulation settings of the generator.
[:SOURce]:CORRection:CSET:CATalog?.........................................................................723
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency.................................................... 723
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?....................................... 724
[:SOURce<hw>]:CORRection:CSET:DATA:POWer........................................................... 724
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?.............................................. 724
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe........................725
[:SOURce]:CORRection:CSET:DELete............................................................................ 725
[:SOURce<hw>]:CORRection:DEXChange:AFILe:CATalog?.............................................. 726
[:SOURce<hw>]:CORRection:DEXChange:AFILe:EXTension.............................................726
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SELect.................................................. 726
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:COLumn................................727
[:SOURce<hw>]:CORRection:DEXChange:AFILe:SEParator:DECimal................................727
[:SOURce<hw>]:CORRection:DEXChange:EXECute.........................................................728
[:SOURce<hw>]:CORRection:DEXChange:MODE............................................................ 728
[:SOURce<hw>]:CORRection:DEXChange:SELect............................................................729
[:SOURce<hw>]:CORRection:CSET[:SELect]................................................................... 729
[:SOURce<hw>]:CORRection[:STATe]............................................................................. 730
[:SOURce<hw>]:CORRection:VALue?............................................................................. 730
[:SOURce<hw>]:CORRection:ZERoing:STATe................................................................. 730
[:SOURce]:CORRection:CSET:CATalog?
Requests a list of user correction tables. The individual lists are separated by commas.
The lists are stored with the fixed file extensions *.uco in a directory of the user's
choice. The directory applicable to the commands is defined with the command
MMEMory:CDIR.
Return values:
<Catalog> string
Example:
MMEM:CDIR '/var/user/ucor'
selects the directory for the user correction files.
CORR:CSET:CAT?
queries which correction tables are available.
Response:UCOR1,UCOR2,UCOR3
the correction tables UCOR1, UCOR2 and UCOR3 are availa-
ble.
Usage: Query only
Manual operation: See "Directory, File List and File Name" on page 161
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency <Frequency>
Transfers the frequency data to the table selected with :CORRection:CSET:SELect.
The numerical suffix at SOURce must not be used for this command.
SOURce Subsystem