EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 285

Rohde & Schwarz SMBV100A
911 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
270Operating Manual 1176.8016.02 ─ 17
To access the pulse generator settings ...
Select "RF/A Mod > config... > Pulse Generator" or use the MENU key under
"RF/A Mod". Alternatively, you can configure the pulse generator in the "Pulse
Modulation" dialog of the "RF/A Mod" block.
The dialog provides the settings for the pulse characteristics and trigger mode.
Depending on the selected modulation source and pulse mode the provided
parameters vary.
Video Sync Signal State - Pulse Generator
Switches on/off the output of the video/sync signal at the PULSE VIDEO connector.
The signal output and the pulse generator are automatically switched on with activation
of pulse modulation if pulse generator is selected as modulation source. The signal
output can be switched off subsequently.
Remote command:
[:SOURce<hw>]:PGENerator:STATe on page 781
Pulse Mode - Pulse Generator
Sets the mode of the pulse generator.
"Single"
A single pulse is generated in one pulse period.
"Double"
Two pulses are generated in one pulse period. Additional settings for
the double pulse are available in the menu.
Remote command:
[:SOURce<hw>]:PULM:MODE on page 799
Pulse Period - Pulse Generator
Sets the period of the generated pulse. The period determines the repetition frequency
of the internal signal.
Remote command:
[:SOURce<hw>]:PULM:PERiod on page 800
RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals