Remote Control Commands
R&S
®
SMBV100A
799Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:PULM:DOUBle:STATe <State>
Activates double pulse generation. The two pulses are generated in one pulse period.
Parameters:
<State> 0 | 1 | OFF | ON
*RST:  0
Example: 
PULM:DOUB:STAT ON
double-pulse mode is enabled.
Options:  R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
Sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width> float
Range:  10 ns  to  100 s
Increment:  10 ns
*RST:  3 us
Example: 
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Options:  R&S SMBV-K23 (Pulse Generator)
Manual operation:  See " Double Pulse Width - Pulse Generator " on page 271
[:SOURce<hw>]:PULM:MODE <Mode>
Sets the mode of the pulse generator.
Parameters:
<Mode> SINGle | DOUBle
SINGle
Enables single pulse generation.
DOUBle
Enables double pulse generation. The two pulses are generated
in one pulse period.
*RST:  SINGle
Example: 
PULM:MODE DOUB
enables double pulse generation.
Options:  R&S SMBV-K23 (Pulse Generator)
Manual operation:  See " Pulse Mode - Pulse Generator " on page 270
[:SOURce<hw>]:PULM:OUTPut:SYNC[:STATe] <Sync>
Configures the signal at the SIGNAL VALID connector.
SOURce Subsystem