Instrument Function
R&S
®
SMBV100A
290Operating Manual 1176.8016.02 ─ 17
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 738
Mode - List Mode
Selects the cycle mode of the List mode.
"Auto"
Cycle from the beginning to the end of the list with automatic restart
at the beginning. If a different mode was activated prior to the Auto
mode, the cycle continues from the beginning of the list. The duration
of a list step is determined by the set dwell time.
Button "Reset" restarts the list at the starting point.
"Single"
Single cycle from the beginning to the end of the list. If "Single" is
selected, the cycle is not started immediately. The "Execute Single"
button appears under the "Mode" line. The cycle is started with this
button. The duration of a list step is determined by the set dwell time.
Button "Reset" restarts the list at the starting point.
"Step"
Manual, step-by-step processing of the list. Activating "Step" stops
the current list and the cursor moves to the value displayed for "Cur-
rent Index". It is now possible to scroll up and down in the list in dis-
crete steps by varying the index. The duration of a list step is deter-
mined by the time between two index entries.
Button "Reset" restarts the list at the starting point.
"Extern Single"
Single cycle from the beginning to the end of the list as with "Single",
but started by an external trigger.
The external trigger signal is input at the BNC connector INST TRIG.
Button "Reset" restarts the list at the starting point.
"Extern Step"
Step-by-step cycle using the external trigger signal. Each trigger
event starts a single step. The duration of a list step is determined by
the time between two trigger events.
The external trigger signal is input at the BNC connector INST TRIG.
Button "Reset" restarts the list at the starting point.
Remote command:
[:SOURce<hw>]:LIST:MODE on page 777
[:SOURce<hw>]:LIST:TRIGger:SOURce on page 779
Execute Single - List Mode
Triggers the list manually. This button is available only if mode "Single" is selected.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:EXECute on page 779
Reset - List Mode
Resets the list to the starting point.
RF Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals