EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 387

Rohde & Schwarz SMBV100A
911 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMBV100A
372Operating Manual 1176.8016.02 ─ 17
The "Trigger In" section is where the trigger for the modulation signal is set. Various
parameters will be provided for the settings, depending on which trigger source - inter-
nal or external - is selected. The current status of signal generation ("Running" or
"Stopped") is indicated for all trigger modes.
The "Marker Mode" section is where the marker signals at the MARKER output con-
nectors are configured.
The "Marker Delay" section is where a marker signal delay can be defined, either with-
out restriction or restricted to the dynamic section, i.e., the section in which it is possi-
ble to make settings without restarting signal and marker generation.
The "Clock Settings" section is where the clock source is selected and - in the case of
an external source - the clock type.
The "Global Trigger/Clock Settings" button leads to a submenu for general trigger,
clock and external input settings.
Trigger In
The "Trigger In" section is used to configure the trigger signal for the digital modulation.
The current status of signal generation is indicated for all trigger modes.
Trigger Mode
Selects trigger mode.
"Auto"
The digital modulation signal is generated continuously.
"Retrigger"
The digital modulation signal is generated continuously. A trigger
event (internal with "Execute Trigger" or external) causes a restart.
"Armed_Auto"
The digital modulation signal is generated only when a trigger event
occurs. Then the digital modulation signal is generated continuously.
Button "Arm" stops signal generation. A subsequent trigger event
(internal with "Execute Trigger" or external) causes a restart.
Baseband Signal - Baseband Block

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals