Instrument Function
R&S
®
SMBV100A
406Operating Manual 1176.8016.02 ─ 17
Offset DC (Rect Test Signal)
Enters a DC component.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:OFFSet on page 616
Generate Signal HD (Rect Test Signal)
Generates a signal and saves it to a file.
The File Select window opens automatically and the signal can be stored as a wave-
form file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:CREate:NAMed
on page 615
Generate Signal RAM (Rect Test Signal)
Generates a signal and uses it as output straight away.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:SINE:CREate on page 615
Const IQ Test Signal
To access the "Sine Test Signals" dialog, in the "ARB MOD" dialog set the parameter
"Create Test Signal" to Const I/Q and press "Create Test Signal".
The dialog can be used to configure a continuous test signal with constant IQ. The I
and Q values for each path are adjustable but constant. The I and Q values are
entered as a decimal number, which is decimal-to-binary converted internally. The sig-
nal is provided as a 16-bit wide digital signal for both I and Q channels.
Converted to analog the signal is output directly at the BNC connectors I and Q outputs
of the instrument.
I Value (Const IQ Test Signal)
Enters the value for the I component of the test signal.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:I on page 615
Q Value (Const IQ Test Signal)
Enters the value for the Q component of the test signal.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:CIQ:Q on page 615
Generate Signal HD
Generates a signal and saves it to a file.
Baseband Signal - Baseband Block