Remote Control Commands
R&S
®
SMBV100A
576Operating Manual 1176.8016.02 ─ 17
Return values:
<Lower> float
Increment: 0.01
Example:
OUTP:AFIX:RANG:LOW?
queries the minimum level for the FIXed setting.
Example:
Response: -50
The minimum level is -50 dBm.
Usage: Query only
Manual operation: See " Fixed Range (PEP) In " on page 224
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set when the attenuator is fixed, see :
OUTPut<hw>:AMODe.
Return values:
<Upper> float
Increment: 0.01
Example:
OUTP:AFIX:RANG:UPP?
queries the maximum level for the FIXed setting for the RF out-
put.
Example:
Response: -27
The maximum level is -27 dBm.
Usage: Query only
Manual operation: See " Fixed Range (PEP) In " on page 224
:OUTPut<hw>:AMODe <AMode>
Selects the mode of the attenuator at the RF output (Attenuator MODe).
Parameters:
<AMode> AUTO | FIXed
AUTO
The electronically switching attenuator switches with a 5 dB step
width at fixed switching points.
FIXed
The attenuator is fixed at the current position. The uninterrupted
level settings are made if automatic level control is activated
(SOURce:POWer:ALC ON).
*RST: AUTO
Example:
POW:ALC ON
activates automatic level control for RF output.
OUTP:AMOD FIX
sets the fixed mode with uninterrupted level for RF output.
OUTPut Subsystem