Remote Control Commands
R&S
®
SMBV100A
665Operating Manual 1176.8016.02 ─ 17
[:SOURce<hw>]:BB:DM:FILTer:OSAMpling:AUTO <Auto>
The command activates/deactivates the upsampling factor state. If activated, the most
sensible parameter values are selected. The value depends on the coherence check. If
deactivated, the values can be changed manually.
Parameters:
<Auto> 0 | 1 | OFF | ON
*RST: 1
Example:
BB:DM:FILT:OSAM:AUTO ON
The most sensible parameters are selected automatically.
Manual operation: See "Oversampling" on page 361
[:SOURce<hw>]:BB:DM:FILTer:PARameter:COSine[:ROLLoff] <Cosine>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:GAUSs <Gauss>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:LPASs <LPass>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:LPASSEVM <LPassEvm>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:PGAuss <PGauss>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:RCOSine <RCosine>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:SPHase <SPhase>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25 <Apco25>
The command sets the roll-off factor for filter type APCO25.
On selecting filter APCO25, the filter parameter is set to the default value.
Parameters:
<Apco25> float
Range: 0.05 to 0.99
Increment: 0.01
*RST: 0.2
<Cosine> float
Range: 0.05 to 1.00
Increment: 0.01
*RST: 0.35
<Gauss> float
Range: 0.15 to 100000
Increment: 0.01
*RST: 0.3
<LPass> float
Range: 0.05 to 2
Increment: 0.01
*RST: 0.5
<LPassEvm> float
Range: 0.05 to 2
Increment: 0.01
*RST: 0.5
SOURce Subsystem