Remote Control Commands
R&S
®
SMBV100A
720Operating Manual 1176.8016.02 ─ 17
Usage: Query only
Manual operation: See "Overload Hold" on page 457
[:SOURce<hw>]:BBIN:OLOad:STATe?
This command queries the current overload state of the A/D converter.
Return values:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:STAT?
queries the overload state.
Response: 0
The AC/DC converter is not overloaded.
Usage: Query only
Manual operation: See "Overload" on page 457
[:SOURce<hw>]:BBIN:POWer:PEAK <Peak>
This command enters the peak level of the external baseband signal relative to full
scale of 0.5 V (in terms of dB full scale).
Parameters:
<Peak> float
Range: -60 to 3.02
Increment: 0.01
*RST: 0
Default unit: dBfs
Example:
BBIN:POW:PEAK -2
enters the peak level of -2 dBfs.
Manual operation: See "Peak Level" on page 456
[:SOURce<hw>]:BBIN:POWer:RMS?
This command queries the rms level of the external digital baseband signal.
Return values:
<Rms> float
Range: -100 to 10
Increment: 0.01
*RST: 0
Example:
BBIN:POW:RMS
queries the estimated rms level.
SOURce Subsystem