Remote Control Commands
R&S
®
SMBV100A
719Operating Manual 1176.8016.02 ─ 17
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:IQSW ON
swaps the I and Q channel of the external baseband signal.
Manual operation: See "I/Q-Swap" on page 454
[:SOURce<hw>]:BBIN:MPERiod <MPeriod>
Sets the recording duration for measuring the baseband input signal by Auto Level Set.
Parameters:
<MPeriod> integer
Range: 1 to 32
*RST: 2
Default unit: s
Example:
BBIN:MPER 4s
sets the recording duration for Auto Level Set to 4 seconds.
Manual operation: See "Measurement Period" on page 456
[:SOURce<hw>]:BBIN:OLOad:HOLD:RESet
This command executes a reset of the Overload Hold State and LED.
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:RES
resets the overload hold state.
Usage: Event
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe?
This command queries an overload since the last reset for evaluating the measure-
ment. The Overload Hold state can be reset via the "Reset" command, or is reset auto-
matically by starting a new measurement ("Auto Level Set") or by setting new values
(Crest Factor, Peak Level, Level).
Return values:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:STAT?
queries the overload state.
Response: 0
the AC/DC converter has not been overloaded.