Remote Control Commands
R&S
®
SMBV100A
718Operating Manual 1176.8016.02 ─ 17
Parameters:
<CFactor> float
Range: 0 to 30
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BBIN:CFAC 10
enters a crest factor of 10 dB for the external baseband signal.
Manual operation: See "Crest Factor" on page 456
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe <State>
Activates automatic adjustment of the baseband input signal.
The R&S SMBV receives peak level, level and crest factor values directly from the con-
nected transmitter and recognizes changes automatically.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
BBIN:DIG:ASET:STAT ON
switches on automatic adjustment of the baseband input signal.
Manual operation: See "DIG IQ Auto Setting" on page 455
[:SOURce<hw>]:BBIN:FOFFset <FOffset>
Enters the frequency offset for the external baseband signal.
The complex I/Q bandwidth of the shifted useful signal must not exceed the total avail-
able baseband bandwidth (see data sheet).
Parameters:
<FOffset> float
Range: -40E6 to 40E6
Increment: 0.01
*RST: 0
Default unit: Hz
Example:
BBIN:FOFF 2 MHz
sets a frequency offset of 2 MHz.
Manual operation: See "Frequency Offset" on page 457
[:SOURce<hw>]:BBIN:IQSWap[:STATe] <State>
This command swaps the I and Q channel if set to ON.
SOURce Subsystem