EasyManua.ls Logo

Rohde & Schwarz SMBV100A - Page 780

Rohde & Schwarz SMBV100A
911 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100A
765Operating Manual 1176.8016.02 ─ 17
Parameters:
<Spacing> LINear | LOGarithmic
LINear
With the linear sweep, the step width is a fixed frequency value
which is added to the current frequency. The step width for lin-
ear sweep is entered in Hz (see [:SOURce<hw>]:LFOutput:
SWEep[:FREQuency]:STEP[:LINear] on page 765).
LOGarithmic
With the logarithmic sweep, the step width is a constant fraction
of the current frequency. This fraction is added to the current fre-
quency. The logarithmic step width is entered in % (see [:
SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:
LOGarithmic on page 766).
*RST:
LINear
Example:
LFO:SWE:SPAC LIN
selects linear sweep spacing.
Manual operation: See " Spacing " on page 267
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step size for linear LF frequency sweep steps.
This parameter correlates with the number of steps [:SOURce<hw>]:LFOutput:
SWEep[:FREQuency]:POINts within the sweep range as follows:
f
STARt
< f
STOP
freq_points = ((f
STARt
- f
STOP
) / step_lin) + 1
If you change the step size, the number of steps changes accordingly. The sweep
range remains the same.
Parameters:
<Linear> float
Range: full frequency range
Increment: see the data sheet: Modulation sources > Resolu-
tion of frequency setting
*RST: 1000
Example:
LFO:FREQ:STAR
sets the start frequency to 2 kHz.
LFO:FREQ:STOP
sets the stop frequency to 20 kHz.
LFO:SWE:SPAC LIN
sets linear sweep spacing.
LFO:SWE:STEP 2 kHz
sets the sweep step width to 2 kHz. The number of sweep steps
for linear sweep spacing (POINts) is automatically set to 11.
Manual operation: See " Step Lin/Log - LF Sweep " on page 268
SOURce Subsystem

Table of Contents

Other manuals for Rohde & Schwarz SMBV100A

Related product manuals