Remote Control Commands
R&S
®
SMBV100A
808Operating Manual 1176.8016.02 ─ 17
The keyword [:FREQuency] can be omitted (see example). The command is then
SCPI-compliant.
Tip: It is recommended to switch off the "Display Update" for optimum sweep perform-
ance especially with short dwell times (SYSTem:DISPlay:UPDate OFF).
Parameters:
<Dwell> float
Range: 2E-3 to 100
Increment: 100E-6
*RST: 15E-3
Example:
SWE:DWEL 12 ms
sets a dwell time of 12 ms for a frequency sweep at the RF out-
put.
Manual operation: See " Dwell Time - Frequency Sweep " on page 282
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute
Starts an RF frequency sweep cycle manually.
The command is only effective in single mode.
Example:
TRIG:FSW:SOUR SING
SOUR:SWE:FREQ:MODE AUT
SWE:FREQ:EXEC
triggers a frequency sweep at the RF output.
Usage: Event
Manual operation: See " Execute Single Sweep - Frequency Sweep " on page 279
[:SOURce<hw>]:SWEep[:FREQuency]:MODE <Mode>
Sets the sweep mode.
The keyword [:FREQuency] can be omitted (see example). The command is then
SCPI-compliant.