Remote Control Commands
R&S
®
SMBV100A
833Operating Manual 1176.8016.02 ─ 17
Example:
SYSTem:COMMunicate:SERial:BAUD 115200
Sets 115200 baudrate.
Manual operation: See "Baud Rate" on page 144
:SYSTem:COMMunicate:SERial:PARity <Parity>
Sets the parity for the serial remote control interface.
Parameters:
<Parity> NONE | ODD | EVEN
*RST: n.a. (factory preset: NONE)
Example:
SYST:COMM:SER:PAR NONE
Selects parity NONE.
Manual operation: See "Parity" on page 144
:SYSTem:COMMunicate:SERial:SBITs <SBits>
Sets the number of stop bits for the serial remote control interface.
Parameters:
<SBits> 1 | 2
*RST: n.a. (factory preset: 1)
Example:
SYST:COMM:SER:SBIT 2
Selects 2 stop bits.
Manual operation: See "Stop Bits" on page 144
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:SOCKet:RESource?
Response: "TCPIP::10.113.1.150::5025::SOCKET"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 144
:SYSTem:FILes:TEMPorary:DELete
Deletes the temporary files from the internal memory or, if installed, from the hard disk.
Example:
:SYSTem:FILes:TEMPorary:DELete
All temporary files are removed.
Usage: Event