Remote Control Commands
R&S
®
SMBV100A
834Operating Manual 1176.8016.02 ─ 17
:SYSTem:MMEMory:HDD?
Queries the status of the mass memory hard disk drive.
Return values:
<State> 0 | 1 | OFF | ON
*RST: ON
Example:
SYST:MMEM:HDD?
Response: 1
the hard disk drive is ready for use.
Usage: Query only
:SYSTem:IDENtification <Identification>
Selects the mode the instrument identification is performed.
Parameters:
<Identification> AUTO | USER
AUTO
The *IDN string and the *OPT string are set automatically.
USER
Enables the selection of user definable *IDN and *OPT strings.
*RST: n.a. (factory preset: AUTO)
Example:
SYST:IDEN USER
Selects the user-defined identification string.
Manual operation: See "Mode" on page 145
:SYSTem:IDENtification:PRESet
Sets the *IDN and *OPT strings in user defined mode to default values.
Example:
SYSTem:IDENtification USER
SYSTem:IDENtification:PRESet
Usage: Event
Manual operation: See "Set to default" on page 146
:SYSTem:IRESponse <IdnResponse>
Defines the user defined identification string for *IDN.
Note: While working in a emulation mode, the instrument's specific command set is
disabled, i.e. the SCPI command SYST:IRES is discarded.