Remote Control Commands
R&S
®
SMBV100A
842Operating Manual 1176.8016.02 ─ 17
Parameters:
<Clock> integer
Range: 100 to 1E8
*RST: 1E6
Manual operation: See "Clock Frequency" on page 205
:TEST:BB:DATA:OFFTime <OnTime>
:TEST:BB:DATA:ONTime <OnTime>
Sets the data enable time interval of the BER/BLER test generator.
Parameters:
<OnTime> integer
Range: 0 to 4294967295
*RST: 1000
Manual operation: See "On Time, Off Time" on page 205
:TEST:BB:DATA:RDELay
<RestartDelay>
Sets the delay time for the restart signal of the BER/BLER test generator.
Parameters:
<RestartDelay> integer
Range: 0 to 4294967295
*RST: 1000
Manual operation: See "Restart Delay" on page 206
:TEST:BB:DATA:STATe <State>
Activates the test generator for the bit error rate measurement, or the block error rate
measurement respectively.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Manual operation: See "State" on page 205
:TEST:BB:DATA:TRIGger[:MODE] <TriggerMode>
Selects the trigger input mode for the BER/BLER test generator.
Parameters:
<TriggerMode> NONE | DENable | RESTart
*RST: NONE
Manual operation: See "Trigger" on page 205
TEST Subsystem