Sirona Dental Systems GmbH 5 Care, cleaning and maintenance by the practice team
Operating Instructions SINIUS / SINIUS CS 5.6 Sanitation
båÖäáëÜ
2. If you want to exclude the instruments from sanitation, touch the
Exclude instrument
key.
Connected instruments will not be rinsed in sanitation phase 5.
Sanitation phase 5 – Rinsing water lines with water
The DENTOSEPT P disinfectant is rinsed out of the SPRAYVIT hoses,
instrument hoses, and tumbler filling unit with water. This takes several
minutes.
Sanitation phase 6 – Returning instruments to their holders
After the water lines have been rinsed, the instruments and suction hoses
can be removed from the receptacle on the water unit and re-inserted.
The
Return instruments to their holders
display appears on the
touchscreen.
Place the SPRAYVIT sleeves, straight and contra-angle handpieces
and suction cannulas back on the instruments and suction hoses. Set
down all instruments and suction hoses again.
The sanitation process is finished. The treatment center is again
ready for operation.
Display in case of incomplete sanitation
The sanitation program detects whether the sanitation of all instruments
and the tumbler filling was fully completed. Errors that occurred during
sanitation will be displayed on the touchscreen as follows:
Crossed-out, solid gray circle: The instrument was excluded from
sanitation before it had been filled with DENTOSEPT P disinfectant
Orange circle with warning triangle: The instrument or tumbler filling
was not (sufficiently) rinsed; the water lines still contain disinfectant
In the latter case, rinse the affected instruments and tumbler filling
manually after sanitation.
Aborting the sanitation process
Sanitation can be canceled during some sanitation phases, e.g. if the
treatment center urgently must be switched back on before the 24 hours
have expired. The
Stop
key shows whether cancellation is possible.
The sanitation process must have been started.
Touch the
Stop
key on the touchscreen.
The sanitation procedure is canceled. If the water lines have not
yet been treated with DENTOSEPT P disinfectant (prior to
sanitation phase 3), the treatment center automatically proceeds
directly to sanitation phase 6. If the sanitation process is
canceled during the reaction time, the water lines are initially
rinsed with water (sanitation phase 5).