5 Care, cleaning and maintenance by the practice team Sirona Dental Systems GmbH
5.6 Sanitation Operating Instructions SINIUS / SINIUS CS
Sanitation phase 3 – Treating water lines with DENTOSEPT P
disinfectant
The water is pumped out of the water tank of the water unit via the tumbler
filler. Then the water tank is automatically filled with DENTOSEPT P. This
activates flushing of the instrument hoses, the SPRAYVIT hoses and the
tumbler filler with the DENTOSEPT P disinfectant. The suction hoses are
cleaned.
Sanitation phase 4 – Allow 24 hours reaction time
In order to effectively combat germs, you must let the DENTOSEPTP
disinfectant react for at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
✔ Sanitation phase 4 is highlighted on the touchscreen.
✔ The treatment center displays the remaining reaction time on the
touchscreen next to the sanitation phase 4 field, starting from 24
hours.
✔ The treatment center has automatically switched to the Standby
mode
1. Leave all instruments plugged into their water unit.
2. Make sure that the treatment center is switched off for at least
24 hours, but not longer than 3 days (maximum sanitation time). If
necessary, you can also switch off the power switch on the base of
the treatment center patient chair.
3. Switch the treatment center on again after 24 hours and before 3
days have elapsed.
Error message: Inserting instruments in the water unit
If an instrument is accidentally deposited in its holder during the reaction
time, the message
Insert instruments into water unit
is displayed on the
touchscreen after the treatment center is switched on. The sanitation
program is then completed anyway.
1. Remove the accidentally removed instrument and place it back in the
receptacle on the water unit so that it can be rinsed in sanitation
phase 5.
IMPORTANT
Switched-off water and air supplies
If the treatment center is switched on again after 24 hours, the sanitation
process automatically continues with sanitation phase 5. However, if the
water and air supplies are switched off, the DENTOSEPT P disinfectant
cannot be rinsed out of the water lines.
If you switch the treatment center back on after 24 hours have expired,
first check to make sure that the water and air supply lines are open.