8.
(For calibrations that pass) Select an option.
Option Description
Details View the calibration images.
Transfer EDS Transfer the calibration file to Connect, a USB drive, or a network drive.
Reject results The calibration file is deleted.
Accept results The calibration file is saved.
If the results are accepted and the calibration file is saved, previous calibration files associated with
the combination of the instrument and the block will no longer be used.
9.
(For calibrations that fail) Select an option.
Option Description
Details View the calibration images.
Transfer EDS Transfer the calibration file t
o Connect, a USB drive, or a network drive (recommended
for troubleshooting).
Close The calibration file is deleted.
See “Troubleshoot calibration failure” on page 95.
10.
Unload the T
aqMan
Array Card from the instrument.
CAUTION! PHY
SICAL INJURY HAZARD. During instrument operation, the TaqMan
Array
Card temperature can reach 100°C. Allow it to cool to room temperature before handling.
Note: Use Smar
t Help if the instrument does not eject the array card (see “Request technical
support with Smart Help” on page 172).
View the calibration images
The instrument performs the ROI, uniformity, and background calibrations in sequence. You can view
the calibr
ation images after the background calibration is complete.
1.
In the home screen, tap
(Settings)4Maint
enance and service4Calibration4Status and
reminders.
The Status and Reminders screen is displayed.
2.
Tap a calibration type in the table, then tap View.
3.
Tap a calibration date in the table, then tap View.
The following details are displayed:
Chapter 7 C
alibrate and verify instrument performance
Perform ROI/uniformity, background, dye, and normalization calibrations for the TaqMan
Array Card block
7
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide
101