4.
Enter a back up file name.
5.
Select the elements to back up.
•
Instrument settings
•
User profiles
•
Calibration records
6.
Tap Back up.
Instrument back up is complete is displayed on the Back Up Instrument screen when the back
up is complete.
7.
Tap Done.
Restore a backup of the instrument
IMPORTANT! A back-up of the instrument includes instrument pr
ofiles and settings. It does not
include plate files or data files. Plate files and data files must be transferred separately. See Chapter 6,
“View and manage files”.
Restoring a backup of the instrument can only be performed by an administrator.
S
ystem templates are included when a backup of the instrument is restored.
1.
In the home screen, tap
(Settings)4Maint
enance and service4Back up/Restore4Restore
a backup.
The Restore a Backup screen is displayed.
2.
Select the backup to restore.
3.
Tap Restore.
Your instrument has been restored is displayed on the Restore a Backup screen.
4.
Tap Done.
Upload a cer
tificate for a TaqMan
™
Array Plate with RFID
1.
In the home screen, tap
(Settings)4Maint
enance and service4RFID certificate.
The TaqMan
™
Array Plate with RFID Certificate screen is displayed.
2.
Tap Upload.
3.
Tap the location of the certificate.
•
Connect
Not
e: You must be signed in with a Connect profile.
•
Network
•
USB
Chapter 11 Configure the instrument settings
Configure maintenance and service settings
11
182
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide