Prepare a calibration TaqMan
Array Card
Required materials to prepare calibration TaqMan
Array Card
Unless otherwise indicated, all materials are available through thermofisher.com. "MLS" indicates that
the material is available from fisherscientific.com or another major laboratory supplier.
Item Source
Calibration kit
ViiA
7 Array Card Spectral Calibration Kit, includes the following
components:
Empty TaqMan
Array Cards
FAM
dye
VIC
dye
ROX
dye
ROI dye mix
Background buer
FAM
/ROX
dye
VIC
/ROX
dye
4432314
Equipment and consumables
Centrifuge with custom buckets and card holders, one of the following:
Sorvall
centrifuge
Multifuge
centrifuge
Megafuge
centrifuge
Contact your local sales oce
TaqMan
Array Card Sealer Contact your local sales oce
Blank balance TaqMan
Array Cards Contact your local sales oce
Pipette, 200 µL with tips MLS
Permanent marker or pen MLS
Powderfree gloves MLS
Safety glasses MLS
Fill the calibration TaqMan
Arr
ay Cards
For detailed instructions to prepare a TaqMan
Array Card, see Appendix C, “Detailed procedures for
preparation of a TaqMan
Array Card”.
IMPORTANT! W
ear powder-free gloves while creating the calibration array cards.
Chapter 7 Calibrate and verify instrument performance
Perform ROI/uniformity, background, dye, and normalization calibrations for the TaqMan
Array Card block
7
98
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide