EasyManua.ls Logo

Thermo Scientific Applied Biosystems QuantStudio 7 Pro - Calibration and verification schedule

Thermo Scientific Applied Biosystems QuantStudio 7 Pro
241 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration and verification schedule
The instrument and block will be calibrated during the installation. To ensure optimal performance,
perform calibrations at the recommended frequency.
Note: After instrument calibration, we recommend performing instrument verification using the
provided RNase P plate.
IMPORTANT! P
erform calibrations and instrument runs under the specified environmental conditions
(see “Environmental requirements” on page 192). Exposure to extreme temperatures can adversely
aect the instrument performance and shorten the life span of the instrument components.
To set calibration frequencies and reminders, see one of the following sections:
V
iew the calibration status on the instrument (see page 87)
View calibration status and set reminders on the Connect platform (see page 88)
Calibration Recommended frequency
ROI/Uniformity
E
very two years (recommended).
Always perform new Background and Dye calibrations after an ROI/Uniformity
calibration.
You will be prompted to run a Background calibration after the ROI/Uniformity
calibration is complete.
The ROI/Uniformity calibration and the Background calibration must pass in order
to accept the results. It is recommended to transfer the files if the calibration fails
in order to troubleshoot.
Note: Performing an ROI/Uniformity calibration invalidates all other calibrations.
Background
Every month (recommended).
Background calibration can also be performed, as needed:
To check for contamination (depends on usage and laboratory conditions).
To obtain the most accurate data for the removal of background
fluorescence.
Note: Performing a Background calibration does not invalidate any other calibration.
Dye
Every two years (recommended).
During a Dye calibration, only the dyes on the given spectral calibration plate are
calibrated.
Note: Performing a Dye calibration for a given dye plate does not invalidate any other
calibration.
Normalization
(TaqMan
Array Card block
only)
Every two years (recommended).
Chapter 7 Calibrate and verify instrument performance
Calibration and verification schedule
7
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide
85

Table of Contents

Related product manuals