Calibrate and verify instrument
performance
Instrument and block calibration ........................................................ 84
Calibration and verification schedule .................................................... 85
Calibration descriptions ............................................................... 86
View the calibration status and set reminders ............................................ 87
Export a calibration report ............................................................. 89
Perform ROI/uniformity, background, and dye calibrations for plate blocks .................. 89
Perform ROI/uniformity, background, dye, and normalization calibrations for the
TaqMan
Array Card block ............................................................ 97
Perform verification using RNase P plates .............................................. 103
Perform verification using RNase P with a TaqMan
Array Card ........................... 108
Calibrate custom dyes ............................................................... 113
Calibrate for a custom melt curve run .................................................. 120
Instrument and block calibration
The QuantStudio
6 Pro System and the QuantStudio
7 Pro System have interchangeable blocks. The
calibration status applies to the combination of the instrument and the block.
If a block is moved to an instrument that it has not been calibrated on, that combination of block and
instrument needs to be calibrated. This applies even if the block has previously been calibrated on a
dierent instrument.
If a block is moved back to an instrument that is has previously been calibrated on, it does not need to
be calibrated again. It needs to be calibrated again if the calibration is expired.
A run can still be started if a calibration of the block on the instrument is expired. A run cannot be
started if a block has never been calibrated on the instrument.
7
84
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide