Detailed procedures for preparation
of a TaqMan
Array Card
Guidelines for preparation of a card
Keep the card protected from light and stored as indicated until ready for use. Excessive exposure
to light may aect the fluorescent probes of the dried-down assays in the card.
Before removing the card from its packaging:
Prepare each sample-specific PCR reaction mix.
Allow the card to reach room temperature.
Load each fill reservoir with 100 µL of sample-specific PCR reaction mix.
Each fill reservoir contains a single sample as determined by the card layout.
The 100–µL volume ensures adequate filling of each reaction well. Volumes smaller than
100 µL result in insuciently filled cards.
Do not allow the micropipette tip to contact the coated foil beneath the fill port.
C
202
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide