Functions that can be audited
Certain instrument functions can be audited. This depends on how the SAE administrator has
configured the audit settings.
The following functions can be audited:
Changes to a plate file, including the following items:
Plate layout
Analysis settings
Run method
Samples
Assays
Sign in success and sign in failure
Sign out
Enable and disable the Security, Auditing, and Esignature (SAE) v2.0 module
Perform a calibration run
Perform an RNase P verification run
Start and stop an instrument run
Pause and resume an instrument run
Functions that can be signed
Certain instrument functions can be signed. This depends on how the SAE administrator has configured
the e-signature settings.
The following functions can be signed with an e-signature:
Start an instrument run
Plate setup
Plate results
Security, Auditing, and Esignature (SAE) v2.0 module roles
Role name
Description
Administrator Has all the available privileges.
Scientist Has all the privileges, except for privileges related to instrument
configuration and maintenance.
Technician Has access to the instrument to perform a run.
Chapter 9 Use the instrument with the Security, Auditing, and Esignature (SAE) v2.0 module
Security, Auditing, and Esignature (SAE) v2.0 module roles
9
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide