(continued)
Calibration description and purpose Pass criteria
Normalization (TaqMan
™
Array Card block only)
•
The software extracts a spectral profile for each normalization standard.
•
The software uses the normalization standards to obtain a correction
factor for the specific dyes, for example, FAM
™
dye or VIC
™
dye.
•
The correction factor is used to generate comparable component signal
levels across instruments.
Normalization standards have
sucient signals to generate a
correction factor.
View the calibration status and set reminders
View the calibration status on the instrument
The Status and Reminders screen also allows you to perform the following functions:
•
Transfer the calibration results (see page 94)
•
Transfer the calibration data files (see page 94)
•
View the calibration images (see page 93)
1.
In the home screen, tap (Settings)4Maintenance and service4Calibration4Status and
reminders.
The Status and Reminders screen is displayed. It shows the following information:
•
Type of calibration
Note:
·
For dye calibration, the type of calibration is listed as the dye that was calibrated, for
example, FAM.
·
Normalization calibration is required for a TaqMan
™
Array Card block. The type of calibration
is listed as FAM/ROX or VIC/ROX.
•
Status of the calibration
•
Last calibration
•
Date that the calibration expires
•
If a reminder is set
2.
Tap a calibration row, then tap View to view the history of that specific calibration type.
•
Dat
e performed
•
Result of the calibration
•
Whether the calibration is current
•
Operator
3.
T
ap Cancel to return to the Status and Reminders screen.
Chapter 7 C
alibrate and verify instrument performance
View the calibration status and set reminders