Review calibration status
The calibration status can be reviewed in the QuantStudio
Design and Analysis Software v2. The
instrument must be added to the Manage Instruments list.
1.
In the home screen, click Manage Instruments.
To access instruments from a dierent screen, click System4Instruments.
2.
In the Summary tab for the instrument, the calibration status is displayed in the Calibration
Status column.
Export a calibration report
The calibration report is exported as a TXT file.
1.
In the home screen, tap (Settings)4Maint
enance and service4Calibration4Status and
reminders.
2.
Tap Export, then select a destination.
Connect
Network Drive
USB Drive
The connectivity status of each location is displayed.
3.
T
ap Done.
Perform ROI/uniformity, background, and dye calibrations
for plate blocks
To calibrate the TaqMan
Array Card block, see “Perform ROI/uniformity, background, dye, and
normalization calibrations for the TaqMan
Array Card block” on page 97.
Chapter 7 C
alibrate and verify instrument performance
Export a calibration report
7
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide
89