Review calibration status
The calibration status can be reviewed in the QuantStudio
™
Design and Analysis Software v2. The
instrument must be added to the Manage Instruments list.
1.
In the home screen, click Manage Instruments.
To access instruments from a dierent screen, click System4Instruments.
2.
In the Summary tab for the instrument, the calibration status is displayed in the Calibration
Status column.
Export a calibration report
The calibration report is exported as a TXT file.
1.
In the home screen, tap (Settings)4Maint
enance and service4Calibration4Status and
reminders.
2.
Tap Export, then select a destination.
•
Connect
•
Network Drive
•
USB Drive
The connectivity status of each location is displayed.
3.
T
ap Done.
Perform ROI/uniformity, background, and dye calibrations
for plate blocks
To calibrate the TaqMan
™
Array Card block, see “Perform ROI/uniformity, background, dye, and
normalization calibrations for the TaqMan
™
Array Card block” on page 97.
Chapter 7 C
alibrate and verify instrument performance
Export a calibration report
7
QuantStudio
™
6 Pro Real-Time PCR System and QuantStudio
™
7 Pro Real-Time PCR System User Guide
89