Set calibration frequency and reminders on the instrument
1.
In the home screen, tap (Settings)4Maintenance and service4Calibration4Status and
reminders.
The Status and Reminders screen is displayed.
2.
Tap a calibration row, then tap Edit.
3.
Tap on the Expiration Interval field, then enter the number of days.
The default is 730 days (2 years).
4.
Tap on the Reminder field, then enter the number of days
The Next Reminder field will be populated.
5.
Set the On/O slider to the On position to enable a reminder about calibration.
6.
Tap Save.
The calibration reminder can be set as a home screen notification (see “Enable and disable home
screen notifications” on page 186).
View calibration status and set reminders on the Connect platform
Note: The calibr
ation reminders feature requires a connection between the instrument and a computer
network.
1.
Sign in t
o Connect on a browser.
2.
Tap
(Instrument) t
o open InstrumentConnect.
3.
Select any of your registered instruments.
4.
In the Summary tab:
Tap
C
alibrations to view the status of each calibration type.
(Optional) Tap
C
alibration Reminders to set the calibration reminder time table and enter
the notification email address or addresses.
Note: More than one email address can receive the calibration reminders.
Note: The settings ar
e automatically saved.
(Optional) In the Downloads section, tap Maint
enance Summary.pdf to download the
calibration status report.
5.
In the Calibrations History tab:
View the history of each calibration type.
(Optional) Tap
to download the calibration history report.
Chapter 7 Calibrate and verify instrument performance
View the calibration status and set reminders
7
88
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide