Custom dye calibration workflow
Determine the optimal dye concentration for each custom dye. Use this concentration to prepare all
subsequent dye calibration plates.
Dilute the custom dye
Prepare a custom dye dilution plate (page 115)
Run the dilution plate (page 116)
Determine the optimal dye concentration (page 116)
Calibrate the custom dye using the optimal concentration
Create a custom dye calibration plate (page 117)
Add a custom dye to the instrument (page 118)
Perform a custom dye calibration (page 118)
Use a dilution series to determine an optimal custom dye concentration
Custom dye dilution guidelines
Prepare a dilution series for each custom dye.
Target several dye concentrations within a range of 100–2,000 nM.
The custom dye for use in calibration must be attached to the 5' end of a short DNA
oligonucleotide consisting of the first two bases of the probe sequence without a quencher at
the 3' end.
Choose a 2 or 3fold dierence in dilution points.
Dispense 10–20 µL per well for 96well plates.
Dispense 5 µL per well for 384-well plates.
Dilute the dye in buer compatible with your Master Mix.
(Intercalating dyes only) Add the appropriate amount of amplified PCR product to generate
fluorescence.
Chapter 7 Calibrate and verify instrument performance
Calibrate custom dyes
7
114
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide