Prepare an RNase P TaqMan
Array Card
For detailed instructions to prepare a TaqMan
Array Card, see Appendix C, “Detailed procedures for
preparation of a TaqMan
Array Card”.
IMPORTANT! W
ear powder-free gloves while preparing the array card.
1.
Remove the V
iiA
7 Array Card RNaseP Verification Kit from the freezer, then allow it to warm to
room temperature (for approximately 15 minutes).
IMPORTANT! You must fill and run the RNase P array card within 60 minutes of thawing the kit.
2.
Remove an arr
ay card from its box and place it on a clean, dry surface.
3.
Using a permanent marker, mark the side of the empty array card with RNase P.
1
1
Location to mark each TaqMan
Arr
ay Card
Chapter 7 C
alibrate and verify instrument performance
Perform verification using RNase P with a TaqMan
Array Card
7
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide
111